The laboratory must calculate its LOD using the standard methodology outlined in ISO/TS 13530:2009 Section 4.4.3. The mathematical calculation must incorporate every single sample preparation, processing, extraction, and analytical step specified in the complete testing method. Because background matrix interference varies, the LOD must be calculated as a direct function of the specific sample type being analysed.